Field vs Aperture Diaphragm Simulator
A microscope has two irises that look alike and do completely different jobs. The field diaphragm sets how much of the specimen is lit. The condenser aperture sets the angle of the light. Set each one and watch the side view, the eyepiece view, and the back focal plane.
Controls
Side view: the condenser lighting the specimen
In the eyepiece
Eyepiece removed
How to set them at the bench
Field diaphragm: close it until you see the leaves in the eyepiece, focus the condenser until their edges are sharp, center them, then open it just until the leaves leave the field of view. Every time you change objectives, check it again, because the field of view gets smaller at higher magnification while the lit area stays the same.
Condenser aperture: take out one eyepiece and look down the tube. Close the aperture until its edge covers about a quarter of the bright circle, so the light fills roughly two thirds to three quarters of the objective's back aperture. That is the usual starting point. Close it more for contrast and depth of field on pale, unstained specimens; open it more when you need every bit of resolution.
Brightness: neither iris is a dimmer. Use the lamp control or neutral density filters for brightness, and use the diaphragms for what they are for.
The numbers here use a green wavelength of 0.55 µm and the Abbe limit, d = λ / (objective NA + illumination NA), with a dry condenser of NA 0.90. A real microscope adds aberrations, coverslip, and alignment, so treat them as a guide. Want to see the same thing from the inside? Ride through it in the Diaphragm Lab or Ride the Light, in 3D or VR. For the full setup, see our Köhler illumination guide.