Field vs Aperture Diaphragm Simulator

A microscope has two irises that look alike and do completely different jobs. The field diaphragm sets how much of the specimen is lit. The condenser aperture sets the angle of the light. Set each one and watch the side view, the eyepiece view, and the back focal plane.

Controls

How much of the specimen is lit
The angle of the cone of light (condenser NA 0.90)
Illumination NA Share of objective NA Finest resolvable detail

Side view: the condenser lighting the specimen

Gold bar: the lit area on the slide, set by the field diaphragm. Rays: their angles come from the condenser aperture (blue) at the bottom. Close one and the other does not change.

In the eyepiece

The field of view, with the image of the field diaphragm leaves on the specimen.

Eyepiece removed

Looking down the tube at the objective's back focal plane. The bright disc is the condenser aperture.

How to set them at the bench

Field diaphragm: close it until you see the leaves in the eyepiece, focus the condenser until their edges are sharp, center them, then open it just until the leaves leave the field of view. Every time you change objectives, check it again, because the field of view gets smaller at higher magnification while the lit area stays the same.

Condenser aperture: take out one eyepiece and look down the tube. Close the aperture until its edge covers about a quarter of the bright circle, so the light fills roughly two thirds to three quarters of the objective's back aperture. That is the usual starting point. Close it more for contrast and depth of field on pale, unstained specimens; open it more when you need every bit of resolution.

Brightness: neither iris is a dimmer. Use the lamp control or neutral density filters for brightness, and use the diaphragms for what they are for.

The numbers here use a green wavelength of 0.55 µm and the Abbe limit, d = λ / (objective NA + illumination NA), with a dry condenser of NA 0.90. A real microscope adds aberrations, coverslip, and alignment, so treat them as a guide. Want to see the same thing from the inside? Ride through it in the Diaphragm Lab or Ride the Light, in 3D or VR. For the full setup, see our Köhler illumination guide.

Need a microscope that sets up the way it should? Sky Optical Services aligns, cleans, and repairs microscopes across Greater Houston.